Model Number 27MJ-501
Device Problems Mechanical Problem (1384); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/19/2021
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2021, a 27mm sjm masters series mechanical heart valve was selected for implant due to calcification.During the procedure, after cardiopulmonary bypass, "one of the leaflets had a problem", however the physician implanted the valve.The patient had problems after the procedure and returned to the operating room, where the valve was exchanged for a new 25mm sjm masters series mechanical heart valve, which was successfully implanted.The physician reported that the surgery was "extremely difficult." no patient consequences were reported.
 
Manufacturer Narrative
Additional information sections: h6, h10 the reported event of "a problem" with one of the leaflets before a smaller valve was implanted could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Manufacturer Narrative
The reported event of a dislodged leaflet was confirmed.One leaflet was dislodged from the orifice.No other damage was noted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The root cause of the leaflet dislodgement could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES MECHANICAL HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key11439976
MDR Text Key238822154
Report Number2648612-2021-00019
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 08/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27MJ-501
Device Catalogue Number27MJ-501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2021
Was the Report Sent to FDA? No
Date Manufacturer Received08/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;