During an atrioventricular node ablation procedure, the generator would not deliver radiofrequency energy and the procedure was cancelled.The generator was reset, new ablation cables were used, and a catheter exchange was attempted, but the issue remained.There were no adverse consequences to the patient.
|
One ampere¿ rf ablation generator was received for evaluation.Visual inspection of the returned generator confirmed all connectors, switches, and labels appeared to have no physical damage.All the mounting hardware was secured.Normal wear from use was observed on the exterior chassis.The engineering evaluation was unable to confirm the field reported event.The returned generator was powered up and the unit completed the post (power on self test) successfully and the screen display came up normally.All connector ports were tested for functionality, numerous catheter codes were manually applied, various impedance & temperature values were also applied for investigation upon which all values were accurately tracked on the generator¿s display screen.Rf energy output was measured during ablation testing and no anomalies were identified throughout investigation as rf output was correctly displayed and measured over an extensive test period.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information provided to abbott and the investigation performed, the reported event was unable to be confirmed.The returned product functioned properly during the evaluation.No hardware abnormalities that would have resulted in the reported event were identified.
|