Model Number CI-1600-04 |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Output Problem (3005); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Inadequate Pain Relief (2388); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 02/03/2021 |
Event Type
Injury
|
Event Description
|
The recipient reportedly experienced poor performance.The recipient's device was explanted.The recipient was reimplanted with another cochlear device.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts to the silastic cover.In addition, the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical test performed.The scanning electron microscopy analysis revealed no evidence of corrosion.This device was explanted for medical reasons.The device passed the tests performed.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|