MDT SOFAMOR DANEK PUERTO RICO MFG PROLOCK CROSSLINK; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
|
Back to Search Results |
|
Model Number G9010001540 |
Device Problems
Break (1069); Difficult to Remove (1528); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a consumer via field representative regarding a patient who undergone spinal therapy with an indication of lumbar spinal canal stenosis (l2/3).It was reported that after the final tightening, it was attempted to remove the crosslink to reinsert screw, but the gold screw of the crosslink fixed to the rod on the right side stripped and it could not remove the crosslink from the rod.The set screw of screw was removed, and the rod was removed together for dealing with it.After reinserting the screw, a new rod and crosslink were placed, and the procedure was completed without problems.There was device breakage and no fragments of the device remained in the patient's body.The device had contact with patient.Tlif was performed at the same level for lumbar spinal stenosis of l2 / 3.No patient symptoms or complications as a result of this event.No additional surgery or treatment performed as a result of this event.There was delay in overall procedure time less than 60 minutes.No in-patient hospitalization or prolongation of existing hospitalization necessary.No further complications were reported.
|
|
Manufacturer Narrative
|
H6: product analysis product id:g9010001452 lot# h5489730 visual and microscopic inspection confirmed the hex of the lock screw has been damaged/stripped.This is consistent with the overload of the set screw during tightening.Additional information: d9, g3, h3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|