(b)(4).Method: the complaint device mr290v vented autofeed humidification chambers were not returned to fisher & paykel healthcare (f&p) in (b)(4) for evaluation.Our investigation is based on the information and photographs provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection of the photographs confirmed holes were found on the aluminium base of the chamber.White-coloured deposits and black-coloured marks were observed on the aluminium base.It was reported by the healthcare facility that 1.4% sodium bicarbonate solutions were unintentionally used with the mr290v vented autofeed humidification chambers.Conclusion: based on the information provided by the customer and previous investigations of similar complaints, the cause of the degradation is due the presence of sodium bicarbonate solution in the mr290v chambers.Sodium salt is highly corrosive to aluminium.The presence of this solution could cause degradation of the aluminium plate over time.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.The subject chamber would have met the required specification prior to distribution.Our user instructions that accompany the mr290v vented autofeed humidification chamber states the following: "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." "use usp sterile water for inhalation or equivalent.".
|