Model Number MR290V
Device Problems Corroded (1131); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint device mr290v vented autofeed humidification chambers were not returned to fisher & paykel healthcare (f&p) in (b)(4) for evaluation.Our investigation is based on the information and photographs provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection of the photographs confirmed holes were found on the aluminium base of the chamber.White-coloured deposits and black-coloured marks were observed on the aluminium base.It was reported by the healthcare facility that 1.4% sodium bicarbonate solutions were unintentionally used with the mr290v vented autofeed humidification chambers.Conclusion: based on the information provided by the customer and previous investigations of similar complaints, the cause of the degradation is due the presence of sodium bicarbonate solution in the mr290v chambers.Sodium salt is highly corrosive to aluminium.The presence of this solution could cause degradation of the aluminium plate over time.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.The subject chamber would have met the required specification prior to distribution.Our user instructions that accompany the mr290v vented autofeed humidification chamber states the following: "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." "use usp sterile water for inhalation or equivalent.".
 
Event Description
A healthcare facility in (b)(6) reported that two mr290v vented autofeed humidification chambers base plates had pin-holes and water was leaking through the pin-holes.The hospital further reported that 1.4% sodium bicarbonate solution was used with the chambers.There were no reported patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key11449313
MDR Text Key244154039
Report Number9611451-2021-00266
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial
Report Date 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? No
Date Manufacturer Received02/08/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-