Catalog Number 71070401
Device Problem Application Program Problem (2880)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/11/2021
Event Type  malfunction  
Event Description
It was reported that software programming issues led to revision surgery.No further information provided.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.Therefore, product analysis could not be performed at this time.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Event Description
It was reported that software programming issue led to a patient revision surgery due to a translated fracture.The malfunction reported happen due to mounting parameter error, the original hardware portion of the device was used.However, a new software program was needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPATIALFRAME.COM SOFTWARE
Type of Device
SOFTWARE FOR DIAGNOSIS/TREATMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11449828
MDR Text Key238810734
Report Number1020279-2021-01932
Device Sequence Number1
Product Code OSN
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71070401
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received03/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKN TAYLOR SPATIAL FRAME EXFIX DEV
-
-