|
Model Number G3603635 |
Device Problems
Material Deformation (2976); Mechanics Altered (2984)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/15/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a consumer via field representative regarding a patient who undergone spinal therapy with an indication of dislocation fracture at c7.It was reported that the screw stripped when the connector was finally tightened.The counter and power grip prevented wiper motion.The screwdriver was firmly inserted in the screw so as not to let the screw stripped.There was no device breakage.Product used correctly according to the directions given in the ifu/labeling.C2-t3 fixation was performed for c7 dislocation fracture.Partial dual rod was performed.Levels implanted c2-t3.No patient symptoms or complications as a result of this event.No in-patient hospitalization or prolongation of existing hospitalization necessary.There was no delay in overall procedure time.No treatment or additional surgery performed as a result of this event.No health damage in the patient was reported.No further complications were reported.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that there was no malfunction reported for any other device and it was confirmed that the allegation "the counter and power grip prevented wiper motion¿ is reported for product ¿g3603635".
|
|
Manufacturer Narrative
|
Product analysis product:3603635 lot# 0719122w visual and optical inspection confirmed the domino is missing one of the set screws.There is no visible damage to the implant.Unable to determine root cause of missing screw.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|