• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE); PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE); PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number UNKNOWN
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hearing Impairment (1881); Infiltration into Tissue (1931); Pain (1994); Metal Related Pathology (4530)
Event Date 10/01/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a revision surgery was performed on the patient´s right hip on (b)(6) 2020.The revision surgery was performed due to chronic pain, hearing loss, and increasingly highly elevated cobalt and chromium levels.The patient outcome is unknown.
 
Manufacturer Narrative
H3, h6: it was reported that right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr head and bhr cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the bhr cup.Similar complaints have been identified for the bhr head.However, as bhr systems of this size are no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.The surgical technique for bhr indicates the acetabular component is to be fully impacted with 15-20° of anteversion and 40-45° inclination angle.The clinical root cause of the reported chronic pain, hearing loss, and elevated cobalt and chromium levels cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate the details supplied in this complaint.The investigation remains inconclusive and a definitive root cause could not be determined.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE)
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key11461626
MDR Text Key239090760
Report Number3005975929-2021-00135
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received06/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-