• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION SD ALARIS SPO2; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SD ALARIS SPO2; PUMP, INFUSION Back to Search Results
Model Number 8220
Device Problems Break (1069); Contamination (1120); Crack (1135); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The affected device has been received and an evaluation is pending.A follow up report will be submitted once the evaluation is completed.Device is pending evaluation.
 
Event Description
The customer reported that the device has a broken case.There was no patient involvement.
 
Event Description
The customer reported that the device has a broken case.There was no patient involvement.
 
Manufacturer Narrative
The investigation is still pending.A follow up report will be submitted once the failure investigation has been completed.
 
Event Description
The customer reported that the device has a broken case.There was no patient involvement.
 
Manufacturer Narrative
This device was evaluated and repaired through the service repair process.Upon visual inspection the service technician noted that they replaced broken rear case, front case, right iui, contaminated patient connector pcb, left iui and lifted keypad.This file is reportable based on the finding of contamination of the device iui.A review of the device history record for (b)(6) was performed, which showed the device had a manufacture date of 15aug2012 and confirmed that this device was not involved in a production failure which correlates to the customer reported issue.The review was performed from the date of manufacture to the date of product release for distribution.A review of the complaint history record was performed for the (b)(6) which did not confirm similar complaints with the same or related failure mode for this customer.
 
Manufacturer Narrative
This device was evaluated and repaired through the service repair process.Upon visual inspection the service technician noted that they replaced broken rear case, front case, right iui, contaminated patient connector pcb, left iui and lifted keypad.The failure code othe was used to track the alaris pump software version as received from the customer and the software version when device was sent back to the customer.It does not reflect a device failure or represent any risk to the patient.A review of the device history record showed the device had a manufacture date of 15aug2012.The review was performed from the date of manufacture to the date of product release for distribution.A review of the device history record in sap for sn (b)(6) was performed which confirmed that this device was not involved in a production failure which correlates to the customer reported issue.A review of the complaint history record in trackwise and sap was performed for the sn (b)(6) which did not confirm similar complaints with the same or related failure mode for this customer.
 
Event Description
The customer reported that the device has a broken case.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SPO2
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key11464284
MDR Text Key239208842
Report Number2016493-2021-29889
Device Sequence Number1
Product Code DQA
UDI-Device Identifier10885403822018
UDI-Public10885403822018
Combination Product (y/n)N
PMA/PMN Number
K010966
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8220
Device Catalogue Number8220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2021
Date Manufacturer Received05/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2720-2020
Patient Sequence Number1
-
-