Model Number V60 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6) 2021.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Event Description
|
The customer called into technical support (ts) reporting that the device is displaying 110b (proximal pressure sensor auto zero failed) error code.The device is not cycling.The customer reported there was no patient involvement at the time the issue was discovered.
|
|
Manufacturer Narrative
|
The remote service engineer provided part id for data acquisition (da) pcba to resolve the reported issue.It is unknown if the customer replaced the data acquisition (da) pcba.Multiple attempts have been made to try to obtain further information regarding the repair and operational status of this case, but no response was received from the customer.
|
|