Model Number V60 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2021.Date of report: 11mar2021.
|
|
Event Description
|
The customer reported there was an error in the touch panel.The field service engineer (fse) confirmed the reported failure.The (fse) calibrated the touch screen; however, the phenomenon did not improve; therefore, the touch screen was replaced.The unit was not in use, and there was no patient or user harm reported.
|
|
Manufacturer Narrative
|
The touchscreen was returned to the manufacturer for evaluation.The customer complaint cannot be verified.After calibration, all buttons respond normally.No-fault found.
|
|