Patient was not receiving adequate ventilation during code blue resuscitation event.It was identified that the ambu bag was missing the white screw in pressure limiting cap.Manufacturer indicated cap is designed to be removable to allow user to clear the valve chamber.The pressure limiting cap should never be removed during operation.Our organization identified ambu packaging can be opened and product manipulated through the package handle.Packaging can be opened/closed and appear unopen.Covid air-borne precaution filters have changed resuscitator product handling for hospitals globally.Filter connections can be confusing (many different products).Opportunity for manufacturer to create an ambu bag that does not have removable piece that makes product inoperable.Other option is to visually indicate piece should not be removed.
|