V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Device Displays Incorrect Message (2591); Failure to Shut Off (2939); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2021.Date of report: 12mar2021.
 
Event Description
The customer reported the unit was connected to a patient in niv.Suddenly the device started alarming and gave an error message "turbine in neutral".After restarting, the screen was frozen and the unit gave another error message: the machine cannot be switched off.The customer contacted product support and requested for service repair.An intervention request was sent to the field team.The customer reported that the unit was in use on a patient, but there was no patient harm reported.The unit was swapped with a working unit.
 
Manufacturer Narrative
G4:12mar2021; b4:30apr2021.The field service engineer (fse) replaced the power management board per field change order to address the reported problem.All checks were carried out according to the manufacturer's recommendations.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11475113
MDR Text Key242646628
Report Number2031642-2021-00946
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Remedial Action Recall
Type of Report Initial,Followup
Report Date 02/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/12/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1622-2020
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER.
-
-