H10: two (2) actual devices were received for evaluation containing fluid in the bladder.Visual inspection was performed which observed vertical cracks located on the fill port.The reported condition was verified for both samples.The cause of the condition could not be determined, however, the probable cause of the cracks may likely be due to the use of excess alcohol solution to clean the fill port either prior, during or after the filling step.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|