ADVANCED BIONICS, LLC CLARION IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100H-11
Device Problems Product Quality Problem (1506); Intermittent Communication Failure (4038)
Patient Problem Tinnitus (2103)
Event Date 03/10/2021
Event Type  malfunction  
Event Description
The recipient reportedly experienced intermittency, sound quality issues, and tinnitus.External equipment was exchanged, however, the issue did not resolve.Device testing revealed results within normal limits.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient's activation reportedly went well.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.The reported complaint of intermittent malfunction could not be verified during this analysis.This device passed all of the electrical tests performed.This older device configuration is no longer manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key11477799
MDR Text Key241480850
Report Number3006556115-2021-00232
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup
Report Date 02/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/01/2003
Device Model NumberAB-5100H-11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age21 YR
-
-