• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1258T/86
Device Problems Signal Artifact/Noise (1036); Low impedance (2285)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2021
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer report number: 2017865-2021-11688.It was reported that the patient presented remotely.Review of the transmission revealed that the left ventricular lead exhibited low out of range impedance and noise reversion episodes and the right ventricular lead exhibited noise, which was recorded as high ventricular rate episodes.Programming changes were made.There were no patient symptoms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11478924
MDR Text Key239624018
Report Number2017865-2021-11687
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734502610
UDI-Public05414734502610
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2017
Device Model Number1258T/86
Device Catalogue Number1258T-86
Device Lot Number4481720
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/23/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TENDRIL ST
Patient Age85 YR
-
-