BARD PERIPHERAL VASCULAR, INC. VACCESS PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number VA80104
Device Problems Retraction Problem (1536); Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 11/2023).
 
Event Description
It was reported that during angioplasty procedure, the balloon allegedly ruptured at 10 atm.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failures as no objective evidence was provided for review.A definitive root cause for the reported balloon rupture and retraction problem could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: b5, d4 (expiry date: 11/2023), g3, h6 (device and method).H11: h6 (conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during angioplasty procedure, the balloon allegedly ruptured at 10 atm.It was further reported that the balloon allegedly had retraction issue.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VACCESS PTA DILATATION CATHETER
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key11482981
MDR Text Key239758367
Report Number2020394-2021-00561
Device Sequence Number1
Product Code LIT
UDI-Device Identifier00801741121258
UDI-Public(01)00801741121258
Combination Product (y/n)N
PMA/PMN Number
K111850
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVA80104
Device Catalogue NumberVA80104
Device Lot Number93YE0015
Was Device Available for Evaluation? No
Date Manufacturer Received07/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-