• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4711K
Device Problem Priming Problem (4040)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the elastomer (bladder) of a folfusor sv (small volume) would not expand enough when introducing an unspecified drug.This was identified prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4 and h6.H4: device manufacture date: august 07, 2020 - august 10, 2020.H10: the actual device was not available; however, a photograph and three (3) companion of the samples were provided for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional flow rate test was performed on the three companion samples and the results were within the product specification range.During and after filling the samples with dextrose solution, no leak was observed at the inlet valve level (fillport) or at other parts of the samples, and the elastomer (bladder) expanded completely without any issues.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Correction: b4/f8: date of this report in mdr follow up #1 is being corrected from blank to 04/15/2021.
 
Manufacturer Narrative
Correction: b4/f8: date of this report in the initial mdr is being corrected from blank to (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11486056
MDR Text Key240342163
Report Number1416980-2021-01328
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412082172
UDI-Public(01)00085412082172
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4711K
Device Lot Number20H038
Was Device Available for Evaluation? No
Date Manufacturer Received07/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NA.
-
-