Catalog Number 112714 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was received for evaluation.A visual inspection observed a crack running in longitudinal direction, starting from the connector borderline and running along a grip.The crack was approximately 19-20 mm length.The lateral area near the thread showed some micro cracks.The reported condition was verified.The cause could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during use with a gamcath, an external leak from the extension tube nearby a connector.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|