Model Number FGS-0635
Device Problems Malposition of Device (2616); Device Dislodged or Dislocated (2923)
Patient Problem Inflammation (1932)
Event Date 02/19/2021
Event Type  malfunction  
Event Description
Incident description: we were attempting to deploy a bravo in the esophagus, the first attempt was unsuccessful: the magnet did not deploy until the probe was removed from the patient, the second attempt: the bravo deployed, but did not attach to the esophagus, it was in the stomach when checked by md.Review: per 0816 am: two endo nursing intra-op minor: primary surgeon description: egd-esophagogastroduodenoscopy with biopsy single or multiple procedure completed: egd-esophagogastroduodenoscopy with biopsy single or multiple.Post-op diagnosis barrett's esophagus.Bravo placement, bravo not deployed.Per endo procedure note by doctor: attempts were made using two bravo ph probes and the probe dislodged both times.One probe was removed and the other one was noted to not have held onto the esophageal mucosa and was loose, so it was pushed into the stomach.The site of attachment where the probe was launched was noted, but it did not hold onto the mucosa.Recommendation: discharge patient to home (ambulatory).Use a proton pump inhibitor po (by mouth) daily.Return to my office in 4 weeks.We are asking the company to evaluate their probes to further evaluate the deployment system.We will need to repeat his exam once the system has been reviewed.Do not use (b)(6), or any other nonsteroidal medications for 7 days.From the procedural note: findings: the z-line was regular and was found 44 cm from the incisors.La grade a (one or more mucosal breaks less than 5 mm, not extending between tops of 2 mucosal folds) esophagitis with no bleeding was found 40 to 43 cm from the incisors.Biopsies were taken with a cold forceps for histology.Estimated blood loss was minimal.The bravo capsule was activated and then calibrated by submersion into the appropriate buffer solutions.The bravo capsule with delivery system was introduced through the mouth and advanced into the esophagus, such that the bravo ph capsule was positioned 38 cm from the incisors, which was 6 cm proximal to the ge (gastroesophageal) junction.The bravo ph capsule was then deployed and attached to the esophageal mucosa.The delivery system was then withdrawn.Endoscopy was utilized for probe placement and diagnostic evaluation.Estimated blood loss was minimal.Attempts were made using 2 bravo ph probes and the probe dislodged both times.One probe was removed and the other one was noted to not have held onto the esophageal mucosa and was loose, so it was pushed into the stomach.The site of attachment where the probe was launched was noted but it did not hold onto the mucosa.The cardia was normal.Localized mild inflammation characterized by erosions and erythema was found in the gastric antrum.The examined duodenum was normal.There were no complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD.
15 hampshire street
mansfield MA 02048
MDR Report Key11526909
MDR Text Key240999445
Report Number11526909
Device Sequence Number1
Product Code FFT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 03/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFGS-0635
Device Catalogue NumberFGS-0635
Device Lot Number51761F
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/12/2021
Event Location Hospital
Date Report to Manufacturer03/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age22630 DA
Patient Weight110
-
-