The device was manufactured from august 17, 2020 - august 18, 2020.The device was evaluated.A visual inspection performed using the naked eye found cracks located on the fill port.The cracks on the fill port resulted in excess residual drug fluid.The cause of the cracks could not be determined; however the likely cause was due to the use of excess alcohol solution to clean the fill port prior, during or after the drug filling step.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|