The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), labeling, applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged guidewire was confirmed.The products returned for evaluation were one 0.038 in.¿j¿-tip guidewire in a plastic hoop and one 18 g introducer needle.The investigation findings are consistent with damage caused by retraction of the guidewire against the bevel of the introducer needle.The returned product sample was evaluated, and the guidewire was confirmed to be broken with the coil wire being unraveled.It appeared that the inner core wire had broken which allowed the outer coil wire surrounding it to unravel.Microscopic examination of the fracture sites revealed the following: narrowing of the wire cross-section near the fracture site, which is a characteristic feature of a strong pull on the wire.Damage to the inside edge of the introducer needle which can occur if the guidewire is forcefully retracted against the needle, damaging the shape of the sharpened bevel.Biological material was also seen on the wire which may have contributed to the observed guidewire fracture as retraction of the wire together with the biological material could have caused the pieces to become stuck.Normal movement of the guidewire is away from the sharpened bevel of the needle and will not damage the wire.If the guidewire direction is reversed, the guidewire is then pulled against the sharpened edge of the needle bevel and can cause shearing damage of the wire and/or cause the wire to become stuck within the needle.An examination of the wire structure revealed no potential damage/defect related to manufacture of the product.The product ifu cautions: ¿do not pull back guidewire over needle bevel as this may sever the end of the guidewire.The introducer needle must be removed first.Also, if unusual resistance is met during manipulation of the guidewire, discontinue the procedure and determine the cause of resistance before proceeding.Withdraw needle and guidewire if cause of resistance cannot be determined.¿ a lot history review (lhr) of rees0476 showed no other similar product complaint(s) from this lot number.
|