ZIMMER BIOMET, INC. COMP PRIMARY STEM 9MM MICRO; EXTREMITY IMPLANT Back to Search Results
Model Number N/A
Device Problems Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cyst(s) (1800); Failure of Implant (1924); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Osteolysis (2377); Subluxation (4525); Metal Related Pathology (4530)
Event Date 11/11/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).Medical product: item number: pt-113950, item name: pt hybrid glen post regenerex, lot number: 819030; item number: 113952, item name: hybrid glenoid base, lot number: 421510; item number: 118001, item name: versa-dial/comp ti std taper, lot number: 076870; item number: 113032, item name: versa-dial humeral head, lot number: 437230.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Reported event was confirmed based on the medical records and x-rays provided which showed tendinitis, mild supraspinatus tendinopathy, possible slight extra-articular impingement of the rotator cuff tendons, osteoarthrosis of the joint, increased periprosthetic bone resorption with thinning of the endosteal surfaces of the left proximal humerus, the humeral articular surface sublux posteriorly relative to its position on prior study.Lucency in the dorsal glenoid could related to granulomatous change in loosening.Findings consistent with loosening of the left shoulder prosthetic device with fracture of the fixating screw in the left glenoid.Intra-articular debris is noted in the left axillary pouch, additional foci of erosions or osteolysis are seen along the inferior aspect of the glenoid which may relate to prior hardware or osteolysis, and possible sebaceous cyst also noted.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-04041.Remains implanted.
 
Event Description
It was further reported that initial left total shoulder arthroplasty performed approximately 3 years ago.Subsequently, the patient reported to her surgeon that she had been doing fine until she reached down to grab something and felt a pain in her left shoulder.Follow up radiology confirmed fracture of the central screw with subsequent loosening and osteolysis from debris generation.The patient reports she has since had pain, difficulty with adls, and limited range of motion.No revision has been reported or planned to date.No further event information available at the time of this report.
 
Event Description
It was further reported that the patient was revised to a reverse total shoulder on approximately a year and a half ago due to glenoid implant fracture.During the revision, evidence of metallosis and wear were found as well as cystic defects and significant scar tissue.All components were removed and replaced with unknown product except for specifying implantation of a biomet base plate.No further information is available at this time and the study is now complete.
 
Manufacturer Narrative
(b)(4).H3: the customer has not indicated whether the product will be returned to zimmer biomet for investigation or not.Once this information is obtained a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-01293, 0001825034-2023-01294 h3 other text : unknown if product will be returned.
 
Manufacturer Narrative
(b)(4).Proposed annex g code: mechanical (g04) - stem.The reported event is confirmed through medical records and mmi review.Review of the available records identified the following: fractured glenoid component (porous metal glenoid post).Revision to reverse total shoulder involving both humeral and glenoid components completed, all study components removed.There was a lot of scarring and obliteration of the tissue planes from the previous surgery.There was metallosis with dark staining of all the soft tissues and synovium from metal-on metal wear.The glenoid component polyethylene face was loose.There was a cystic reaction around the inferior glenoid with a lot of blackened metal wear debris, the joint was meticulously debrided of this.She did have a cystic contained defect from the previous peg holes as well as a large cystic reaction of the inferior glenoid that measured 1.5cm.New definitive components were implanted and excellent stability was achieved.Patient to recovery in good condition with no intra-op complications noted.Review of the radiographs found the following: suspected loosening, and mild osteopenia, glenoid implant appears fractured as well as humeral implant appears larger than expected for patients anatomy.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP PRIMARY STEM 9MM MICRO
Type of Device
EXTREMITY IMPLANT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key11570067
MDR Text Key242411309
Report Number0001825034-2021-00899
Device Sequence Number1
Product Code HSD
UDI-Device Identifier00880304554108
UDI-Public(01)00880304554108(17)261228(10)597750
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060692
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Consumer
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number113609
Device Lot Number597750
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10.
Patient Outcome(s) Other; Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight43 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-