Catalog Number A1059 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported that the locking knob on the mayfield modified skull clamp (a1059) was very loose and considered to be patient safety risk.It is unknown if there was patient involvement; however no patient injury or surgical delay has been reported.
|
|
Manufacturer Narrative
|
Device history record: the dhr shows no abnormalities related to the reported failure.Mayfield modified skull clamp (a1059) was returned for evaluation.The reported complaint was confirmed.The lock and index knob had worn parts.The evaluation showed that the lock had movement and a residue buildup is present.Unit needs to be machined to have heli-coils added to large starburst threads; the set screw in the swivel base was tight.The unit needs repair, and replacement of worn parts.General maintenance and cleaning required.
|
|