• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MANSFIELD TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MANSFIELD TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 72203012
Device Problems Detachment of Device or Device Component (2907); Mechanical Jam (2983); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had an inner blade fracture.It was reported that the device would not oscillate and had an unusual sound.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a diagnostic hysteroscopy procedure, the device was nonfunctional in activating or it stopped rotating.The device was making an unusual noise.The procedure was extended for 5 minutes.A new device was opened and used successfully.There was no patient injury.Medtronic investigation state that there was evidence that the outer blade assembly had been bent at the distal end.Inspection of the inner blade had a deformation on the distal cutting edge.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
Manufacturer Contact
lisa hernandez
5920 longbow drive
boulder, CO 80301
2034925563
MDR Report Key11582213
MDR Text Key242734803
Report Number1282497-2021-00021
Device Sequence Number1
Product Code HIH
UDI-Device Identifier10884521744110
UDI-Public10884521744110
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K130215
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/23/2022
Device Model Number72203012
Device Catalogue Number72203012
Device Lot Number4894642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/04/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age38 YR
-
-