THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, OUS; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 201-10002
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2021
Event Type  malfunction  
Event Description
It was reported that during a training the perfusion team received an m4 alarm on a centrimag motor.The consoles were switched but the alarm persisted and the motor was put aside.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of a m4 alarm was not confirmed as the alarm was not reproduced during testing of the returned centrimag motor (serial number: (b)(6), and no log files were submitted for review.However, damage to the motor that may have contributed to the reported event was confirmed.The returned centrimag motor was evaluated at the european distribution center alongside known working centrimag equipment; however, a m2 and m3 alarm activated after connecting the motor to the console, indicating that the motor was unable to be recognized by the console.The resistance of the wires in the motor cable were measured and it was revealed that two of the motor¿s bearing phase wires were electrically open.The damage to the wires would prevent the motor from functioning properly and may have contributed to the reported event.It was advised to scrap the motor after approval from customer service.The reported event appears to be due to wire fatigue in the motor¿s cable; however, the root cause of the damage to the cable could not be conclusively determined through this analysis.Incidental findings: the screw on the motor cable connector was observed to be loose, damage to motor cable cap.The 2nd generation centrimag system operating manual section 10 ¿ ¿emergencies/troubleshooting¿ provides instructions for operation when there is a need to exchange the main console or motor with a backup console or motor.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.The 2nd generation centrimag system operating manual section 12.1 ¿ "appendix i ¿ primary console alarms and alerts" cover all alarms (auditory and visual), including motor alarms, and the appropriate actions to take if the issue does not resolve.Centrimag motor instructions for use instructs the user to inspect the centrimag motor, cable, console connector, and locking mechanism for any damage prior to use.If any component is damaged, do not use the centrimag motor.This document states that if the unit fails to operate according to the motor specifications or a console diagnostic error indicates a centrimag motor malfunction, it should be returned.Additionally, this document instructs the user to always have a spare centrimag motor and back-up equipment available.The device history records were reviewed and the records revealed that the centrimag motor, serial number (b)(6), was manufactured in accordance with manufacturing and qa specifications.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, OUS
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
MDR Report Key11591120
MDR Text Key253824929
Report Number3003306248-2021-01079
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number201-10002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2021
Was the Report Sent to FDA? No
Date Manufacturer Received06/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberFA-Q318-MCS-1
Patient Sequence Number1
-
-