Catalog Number 76098235
Device Problems Break (1069); Material Erosion (1214)
Patient Problems Failure of Implant (1924); Ambulation Difficulties (2544)
Event Date 05/05/2020
Event Type  Injury  
Event Description
It was reported that the patient presented a fall earlier in the year, with a completely symptom-free interval.After that, the patient had presented a crack sound in the left hip when he was tried to bend down or move.A revision was carried out, where it was possible to observe a fracture of the ceramic head with the destruction of the femoral neck.For this reason, a change of head, the femoral neck and the insert were necessary.The insert was removed because it's necessary to do it the change of head and femoral neck and was also found loose.
 
Manufacturer Narrative
The associated device was returned and evaluated.Visual inspection of retrieved 13mm left image hip stem revealed deformation wear on the proximal male taper and femoral neck, significant scarring and scratches on the neck and around the removal slot, deep gouging on the proximal portion of the stem, and bone on growth on the proximal textured / grooved portion of the stem.The proximal male taper deformation wear is more than likely due to direct contact and repetitive articulation on the acetabular components.Scratches and scarring to the neck region were probably caused by contact with the femoral head debris.Gouging and scarring to the proximal portion of the stem is commonly seen after the revision process, where a sharp instrument is used to remove the implant.Bone on growth indicates that the stem was well fixed in the femoral canal, as intended by design.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.The clinical/medical evaluation concluded that this complaint from germany reports a revision was performed after a fall and cracking noise developed in the left hip.Surgery revealed a fracture of the ceramic head and destruction of the shaft cone.The stem, head and the inlay were changed out.Only the report that the surgery had been done was provided.Smith and nephew has not received adequate materials (operative reports) to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.It is unclear if the fall precipitated the event.Impact to the patient cannot be determined beyond the surgery and recovery.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use document revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.Possible causes could include but not limited to traumatic injury.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMAGE HA HIP SZ 13MM LEFT
Type of Device
PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks road
memphis, TN 38116
5123913905
MDR Report Key11593531
MDR Text Key243092994
Report Number1020279-2021-02565
Device Sequence Number1
Product Code JDH
UDI-Device Identifier03596010389312
UDI-Public3596010389312
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number76098235
Device Lot Number03ET74092
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention; Other;
-
-