DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Gas Output Problem (1266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and was able to reproduce the reported issue.The fse resolved the reported issue by adjusting the internal helium regulator and tightened the helium regulator on the cart.The fse reported that he successfully exercised the unit to no fail.The fse then performed a full calibration, all functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that prior to use, the cardiosave intra-aortic balloon pump (iabp) displayed low helium warnings.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Analysis of production: (3331/213) the device history record review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period apr 2019 through mar 2021 was reviewed.There were no triggers identified for the review period.
|
|
Event Description
|
N/a.
|
|