Qn# (b)(4).The actual device was not returned; however the customer provided a photo for evaluation.A visual exam was performed on the photo and it was observed that the tubing was detached.This may happen due to mishandling of the product during usage.It may also be due to extra force applied to the device during use.In the current manufacturing procedure, 100% leak testing and visual inspection is conducted after the assembly process; thus, defects would be detected prior to release from the manufacturing facility.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.,, if the sample becomes available at a later date a follow up report will be submitted with investigation results.
|