• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. EXTERNAL TEMPORARY PACEMAKER / IMPLANT TOOL MODEL; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. EXTERNAL TEMPORARY PACEMAKER / IMPLANT TOOL MODEL; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5318
Device Problem Impedance Problem (2950)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2021
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis of the external pulse generator (epg) was unable to confirm the customer comments that the impedance measurements were greater than 4000 during testing.Analysis also noted that the bail cover attachment was missing.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generators (epg) impedance was greater than 4000 during testing.The event occurred during set-up.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTERNAL TEMPORARY PACEMAKER / IMPLANT TOOL MODEL
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key11600911
MDR Text Key243311820
Report Number2182208-2021-01282
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K971474
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5318
Device Catalogue Number5318
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2021
Date Manufacturer Received03/02/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-