• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER PUREPOINT LASER; LASER, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER PUREPOINT LASER; LASER, OPHTHALMIC Back to Search Results
Catalog Number 8065750597
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that prior to a procedure the laser display failed.The system turned on but does not present any information on the display.There was no patient involvement.Additional information has been requested.
 
Manufacturer Narrative
The company service representative examined the system and replicated the reported event.The nonconforming printed circuit board (pcb) was replaced to resolve the issue.The system was then tested and met product specifications.The system manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.Root cause the root cause of the reported event can be attributed to the nonconforming printed circuit board (pcb).The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREPOINT LASER
Type of Device
LASER, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key11602971
MDR Text Key244027962
Report Number2028159-2021-00339
Device Sequence Number1
Product Code HQF
UDI-Device Identifier00380657505975
UDI-Public00380657505975
Combination Product (y/n)N
PMA/PMN Number
K062624
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750597
Device Lot Number139C09
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-