Model Number 0035280
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/05/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that foreign material was found inside the syringe.
 
Event Description
It was reported that foreign material was found inside the syringe.Per follow up via ibc on 18mar2021, the issue was found upon opening of the package.
 
Manufacturer Narrative
The reported event was confirmed cause unknown.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.The product was not used for treatment.The product caused the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), used silicone irrigation bulb syringe.Visual inspection of the sample noted no obvious visible defects.It was noted that there was a loose piece of cardboard material on the interior of the syringe barrel.This does not meet product specifications.A potential root cause for this failure could be "no following to production areas cleaning procedure".The dhr review could not be performed without a lot number.A labeling review was not performed because labelling could not have prevented the reported failure.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE
Type of Device
IRRIGATION SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11604206
MDR Text Key243546294
Report Number1018233-2021-01777
Device Sequence Number1
Product Code KYZ
UDI-Device Identifier00801741049255
UDI-Public(01)00801741049255
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 08/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0035280
Device Catalogue Number0035280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2021
Date Manufacturer Received08/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-