MAUDE Adverse Event Report: ALYDIA HEALTH JADA SYSTEM; INTRAUTERINE VACUUM CONTRACTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALYDIA HEALTH JADA SYSTEM; INTRAUTERINE VACUUM CONTRACTION SYSTEM Back to Search Results
Model Number JADA-1001
Device Problem Suction Failure (4039)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 03/03/2021
Event Type  Injury  
Manufacturer Narrative
It is unclear whether the jada malfunctioned based on the information that is available at this time, but out of an abundance of caution and because we cannot rule it out at this time, we are reporting this event due to the 30-day reporting deadline.If we become aware of additional information, we will supplement this report.The submission of this report is not an admission that the device caused or contributed to the reported event.
 
Event Description
It was reported that a (b)(6) patient delivered and experienced a postpartum hemorrhage.Patient was treated with pitocin, hemabate, txa (tranexamic acid) and cytotec.After 1500ml blood loss, the md placed jada and no blood was being drawn through jada.The jada was removed and noted to be clotted.The physician removed any retained products and then troubleshooted by changing the suction tubing and flushing the jada device.There was no suction noted and the device was contaminated due to the patient experiencing diarrhea.The physician used another jada and noted "the same issues of not suctioning" and a bakri balloon was placed.The physician noted clots in the loop of jada upon removal.Total ebl at this time was 2500ml.No further escalation of care was noted.The patient subsequently recovered and was discharged home.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JADA SYSTEM
Type of Device
INTRAUTERINE VACUUM CONTRACTION SYSTEM
Manufacturer (Section D)
ALYDIA HEALTH
3495 edison way
menlo park CA 94025
Manufacturer (Section G)
ALYDIA HEALTH
3495 edison way
menlo park CA 94025
Manufacturer Contact
richard ferrick
3495 edison way
menlo park, CA 94025
8445232666
MDR Report Key11611675
MDR Text Key245973185
Report Number3017425145-2021-00008
Device Sequence Number1
Product Code OQY
UDI-Device Identifier00850017882003
UDI-Public(01)00850017882003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberJADA-1001
Device Catalogue NumberJADA-1001
Device Lot Number1048001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/03/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age38 YR
Patient Weight79
-
-