SMITH & NEPHEW, INC. ACCUR-TIB BS IMPACT CMNT/PRESS; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71236968 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/09/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that, during the journey uni procedure, when using the impactor to place the tibial plate down, the gray portion broke off the impactor.No other complications were reported.
|
|
Manufacturer Narrative
|
H6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical evaluation concluded that this case reports that during a journey uni surgery, a piece of the impactor broke off.Per email correspondence, there was no patient injury or complications, and the procedure was completed with less than a 30-minute delay, using the same device.No further clinical assessment is warranted.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|