• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM INDIA; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM INDIA; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number ROB00036
Device Problem Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2021
Event Type  malfunction  
Event Description
It was reported that during machine setup before a navio procedure, it was found that the orange led indicator on the ups is lit.The ups was kept for charging overnight; however, the battery indicator did not turn green.The batteries need to be replaced.There was no impact on the case.No patient was involved.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the navio surgical system, part rob00036 used for treatment was not made available to the designated complaint unit for evaluation thus, a visual and functional evaluation could not be performed and a relationship between the reported event and the device could not be determined.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed, a factor that may have contributed to the reported event may have been associated with a battery failure.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO SURGICAL SYSTEM INDIA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11612207
MDR Text Key243729320
Report Number3010266064-2021-00246
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556693506
UDI-Public885556693506
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberROB00036
Was Device Available for Evaluation? No
Date Manufacturer Received10/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-