Catalog Number UNK - IMPLANT |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Foreign Body Reaction (1868); Swelling/ Edema (4577)
|
Event Type
Injury
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This report is for an unknown implant.Part and lot number are unknown.Without the specific part number; the udi number, 510-k number and manufacturing site name are unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
Event Description
|
It was reported by the sales rep that a surgeon reported a tibial intrafix device removal from a patient that experienced swelling and irritation at the implant site.The date of the initial surgery was on (b)(6) 2014 but the exact date was unknown.No additional information was provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: according to the information provided, it was reported that the surgeon reported a tibial intrafix device removal from a patient that experienced swelling and irritation at the implant site.The date of the initial surgery was on (b)(6) 2014.The sales rep does not have the product information.The complaint devices were received and evaluated.A screw and sheath of the bio intrafix family devices were returned, however, the product code and the lot number were not provided.Upon visual inspection, it was observed that the screw has rests of biological matter, the screw has no structural anomalies.The sheath has rests of biological matter and is damaged due to the removal procedure.Given that lot number was not provided, the manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection performed, this complaint can be confirmed.The root cause for the reported issue can be attributed to tissue reaction due to the material of the screw and sheath.As stated on ifu 107475: adverse effects of the absorbable implanted devices include mild inflammatory and foreign body reactions.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
|