Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed silicone damage on the top cover, and the electrode was sliced and severed along the lead and near the array prior to receipt.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.The photographic imaging inspection confirmed cut electrode wires along the lead.This is believed to have occurred during revision surgery.System lock was verified.The electrode condition prevented an electrical test performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|