Additional information: d9, h3, h6.Correction to g1 address.H10: the device was received for evaluation.A visual inspection was performed which noted the device had unknown fluids on the display.Functional system level testing was performed with no issues noted.The reported microbial growth was not verified; however, fluids were found on the display.The fluid on the display is outside the fluid path and would not impact the sterility of the solution compounded.The cause of the fluids could not be determined.The display would be serviced and replaced.Should additional relevant information become available, a supplemental report will be submitted.
|