A belmont rapid infuser ri-2 was set up for an or trauma case and the clinical staff noticed flames coming from the back of the device.It was put out with an extinguisher, the fire department called, and the device sequestered.Initial assessment points to the device power plug being the source of failure.Contributing factors may include liquid around the power cord or a loose power connection.Dph has requested that all belmont rapid infusers ri-2 devices be pulled from use until the vendor determines the root cause and clears them for use.Manufacturer response for iv infusion pump, rapid infusor (per site reporter): vendor has been notified and is investigating for root cause.They are suggesting removing all ri-2 devices and replace the fleet with an alternative.
|