• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71953-01
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2021
Event Type  Injury  
Manufacturer Narrative
This is an initial final report. this issue does not meet reportability criteria; however, it is being reported to the fda as the complaint was received via user report.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The device manufacturing date is unknown.The date entered is the date abbott diabetes care became aware of the event.An additional reader 2 (serial numbers unknown) has been captured under this submission.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Abbott diabetes care received a medwatch user report which reported the following information: a customer reported 2 issues he has with the freestyle libre 2 reader settings.The customer stated that the "turning reader off" message displayed when using physical button on the reader is misleading, and leads the customer to believe that the reader will no longer communicate with sensor, instead of locking the touchscreen.As a result, the customer has left the reader "on", but by putting the reader in the pocket in this state, the touchscreen has inadvertently made changes to the settings.Due to this, the customer reported issues with glucose values being recorded under wrong date (date settings changed) and alarm sound being lowered, leading to possible missed alerts.The customer suggested clearer wording when turning touchscreen off, additional information added to device manual, and the ability to correct glucose data if incorrectly recorded.There was no report of any third-party medical intervention due to reported issues and no further details were provided.Based on the information provided, there was no report of serious injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key11634972
MDR Text Key244445553
Report Number2954323-2021-65935
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/08/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71953-01
Was Device Available for Evaluation? No
Date Manufacturer Received03/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/15/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-