As the lot number for the device was not provided, a review of the device history records could not be performed.The device has not been returned to the manufacturer for evaluation.However, image were provided for review.The investigation of the reported event is currently underway.
|
H10: manufacturing review: as the lot number for the device was not provided, a review of the device history records could not be performed.Investigation summary: a device sample or the segment retrieved from the patient were not returned for evaluation.However, an x-ray and photos of the segment were provided for evaluation.An x-ray was provided showing the forearm of a patient, with two stents placed in an arteriovenous fistula.The stents were placed overlapping; a longer stent of a competitor and a shorted lifestent 5f stent.One end of the competitor stent was placed in a curve, while the lifestent 5f was placed in a straight section right after the curve.A segment of a catheter was found inside the lifestent 5f, with an estimated length of 50mm.Two photos were also provided.The first image is showing the distal end of a lifestent 5f delivery system inside a blue kidney basin.The second photo is a close up of the same segment showing the cardan tube of the inner catheter including the atraumatic tip of a lifestent 5f inner catheter.Based on information available and the evaluation of the images provided a break of the inner catheter during the initial stent deployment procedure is confirmed.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the instructions for use supplied with this product the potential issue was found addressed.The instructions for use states: "if resistance is met while retracting the delivery system over a guidewire, remove the delivery system and guidewire together." regarding deployment force the instructions for use states: "do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit." regarding the potential factor of insufficient pre dilation the instructions for use states: "pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' under materials required the instructions for use state: '5f (1.67 mm) or larger introducer sheath (¿) 0.014 inch (0.36 mm) - 0.035 inch (0.89 mm) diameter guidewire." placement of the stent in an av fistula represent an off label use of the device.Based on the instructions for use supplied with this product, the lifestent 5f vascular stent system is intended to improve luminal diameter in the treatment of symptomatic de novo or restenotic lesions in the native superficial femoral artery (sfa) and popliteal artery.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|