EPOCAL INC. EPOC READER Back to Search Results
Model Number HR-1002-00-00
Device Problem Insufficient Information (3190)
Patient Problem Abnormal Blood Gases (1034)
Event Date 03/19/2021
Event Type  malfunction  
Manufacturer Narrative
The customer provided the epoc results they considered discrepant, but no comparative results.The customer's complaint of low po2 results with card lot # 01-21035-20 will be investigated by looking at in-house data, finished goods testing, lifetime testing, retain testing and a search to see if there are any other complaints on this same card lot.The customer stated since they have used a new card package of the same lot, they did not observe any further discrepant results and they are operational.The cause of this event is unknown.
 
Event Description
The customer is alleging discrepant low po2 results on 3 patients on the epoc reader.The customer provided the epoc results they considered discrepant, but no comparative results.The customer stated for elderly ventilated patients, the customer expected values of at least 60 mmhg.For younger patients they expected values of at least 80 mmhg.The customer did not specify which patients are elderly and which are younger.Siemens requested additional information several times but none has been received.The customer stated the samples are collected from the patient's earlobes after the application of finalgon to cause increased blood flow to the area.There is no report of injury due to this event.
 
Manufacturer Narrative
Siemens has completed the investigation: the cause of the alleged discrepancies could not be identified from the information provided by the customer.A review of the in-house performance for the card lot used, lot 01-21035-20, did not identify any product deficiencies.In addition, there were no other complaints received on card lot 01-21035-20.Therefore, there is no further evidence that the system or reagent cards are not performing as intended.A review of finished goods (fg) test data demonstrates that card lot 01-21035-20 performed as expected for po2 with respect to the target.Overall, the lot had acceptable performance for po2 in both aqueous control fluids tested at the time of lot release.The fg blood results for card lot 01-21035-20 were assessed for evidence of discrepant bias on the oxygen sensor.The fg blood results for card 01-21035-20 also demonstrated acceptable performance for po2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPOC READER
Type of Device
EPOC READER
Manufacturer (Section D)
EPOCAL INC.
2060 walkley road
ottawa, ontario K1G 3 P5
CA  K1G 3P5
MDR Report Key11649264
MDR Text Key245386235
Report Number3002637618-2021-00024
Device Sequence Number1
Product Code CGL
UDI-Device Identifier00809708016685
UDI-Public00809708016685
Combination Product (y/n)N
PMA/PMN Number
K113726
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHR-1002-00-00
Device Catalogue Number10736398
Date Manufacturer Received05/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-