SMITH & NEPHEW, INC. SUPINE TABLE ATTACHMENT (A2) W/PAD; APPARATUS, TRACTION, NON-POWERED
|
Back to Search Results |
|
Model Number 72201725 |
Device Problem
Unstable (1667)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/23/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference (b)(4).
|
|
Event Description
|
It was reported that the supine table attachment did not lock away properly during set up for a hip scope.The screw was striped so the attachment could not be securely attached to the bed.There was a smith and nephew back up device.No patient injury, delays or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H3,h6: the reported device, intended for use in treatment was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
|
Manufacturer Narrative
|
H10: internal complaint reference: (b)(4).The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found rust on the sliding housings attached to the bed rails.It was also found that one of the two mast holder knobs was bent.A functional evaluation found one of the two mast holder knobs would not tighten due to the bend in the shaft.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The complaint was verified.Factors that could have contributed to this event include an impact force inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
|