• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 6382
Device Problems Detachment of Device or Device Component (2907); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2021
Event Type  malfunction  
Manufacturer Narrative
The fsr installed a new flow sensor.The unit operated to the manufacturer's specifications.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the flow sensor was missing a pin and was unable to latch properly.There was no patient involvement.
 
Manufacturer Narrative
Updated block: h6 the reported complaint was confirmed.It was determined that the flow sensor was missing a pin that kept it closed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key11671893
MDR Text Key245624702
Report Number1828100-2021-00132
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number6382
Device Catalogue Number6382
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2021
Was the Report Sent to FDA? No
Date Manufacturer Received04/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-