This event occurred in (b)(6).Unique device identifier (udi) (b)(4).The customer confirmed system maintenance was performed regularly.Also, the measuring cell was new, and the pipetting system, mixer unit, pinch valve tubings were controlled and looked good.The patient's sample was requested for an investigation, but the sample was not available.The investigation did not identify a product problem. a general instrument or reagent problem could be identified.The cause of the event could not be determined. .
|
The initial reporter received questionable elecsys insulin results for one patient tested on a cobas e411 rack with a serial number (b)(4).The customer confirmed qc was acceptable.The patient's initial insulin result was not reported outside the laboratory.The customer performed repeat measurements on the same analyzer as well as a cobas 8000 e 801 module located at a different laboratory.On (b)(6) 2021 and at 11:38, the patient's initial insulin result was 0.200 uu/ml with a data flag.At 12:08, the patient's first repeat result on the e411 rack was 0.200 uu/ml with a data flag.At 16:00, the patient's second repeat result on an e 801 module was 1.20 uu/ml.On (b)(6) 2021, the patient's third repeat result on the e411 rack was 0.200 uu/ml with a data flag.The customer reported the patient's insulin result of 1.20 uu/ml outside the laboratory.
|