Manufacturer Narrative
Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.A field service engineering (fse) was at the customer's site to address reported event.While troubleshooting, fse could not confirm or reproduce reported error.Fse performed barcode scan and all 10 barcodes were read, confirming the barcode reader was functioning as expected.The customer then reported data is not crossing over to the lis.Fse checked communication with lis and hyper terminal and got no data.Fse resolved the complaint by replacing the mb board.Fse then checked communication with hyper terminal and got data going to the hyper terminal from the analyzer.Fse successfully validated the analyzer by performing quality control run without errors and within acceptable range.No further action required by field service.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no similar complaints identified during the search period.The most probable cause of the reported event was due to failure of the mb board.
 
Event Description
A customer reported their barcode reader is not reading barcodes on the aia-900 analyzer.The customer stated this is an ongoing issue, and confirms the labels were verified to be good.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for estradiol (e2).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-900
Type of Device
AIA-900
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA   1058623
Manufacturer Contact
bernadette oconnell
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586-23
JA   1058623
MDR Report Key11678698
MDR Text Key249222478
Report Number8031673-2021-00092
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-900
Device Catalogue Number022930111
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-