Model Number N/A
Device Problems Crack (1135); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2020
Event Type  malfunction  
Manufacturer Narrative
The complaint is confirmed for one returned polaris screw for the failure of lacking polyaxial motion, causing mating issues.Medical records were not provided for review.Device evaluation: visual inspection revealed minor scratches on the tulip, but no signs of damage.A functional examination shows that the tulip is frozen and there is no polyaxial movement of the tulip head.The complaint is not confirmed for a crack, however it is confirmed for no polyaxial motion of the tulip head.Potential cause: root cause was unable to be determined.This event could possibly be attributed to forces applied at an off-axis angle.Dhr review and related actions: per dhr review, the part was likely conforming when it left zimmer biomet control.No actions required.This event is not related to any current actions or recalls or product holds.Device use and compatibility: this device is used for treatment.Reported event is not related to a combination of product lines; therefore a compatibility review is not applicable.A follow-up report will be submitted if new information is received that changes the information provided in this report.
 
Event Description
It was reported the polaris screw was found to have a crack at the end of the screw connection during the operation.Another screw was used to complete the operation.No patient impact was reported.However, upon device evaluation by the manufacturer, the screw was found to have lost its polyaxial motion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.5 X 45MM MULTIAXIAL SCREW
Type of Device
EBI 5.5 HELICAL FLANGE SPINAL SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key11680443
MDR Text Key248174893
Report Number3012447612-2021-00122
Device Sequence Number1
Product Code MNH
UDI-Device Identifier00880304925182
UDI-Public(01)00880304925182(10)J6482766
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K061441
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number2000-2445
Device Lot NumberJ6482766
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/15/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age44 YR
Patient Weight63
-
-