A user facility in (b)(6) reported that, during surgery, the surgeon found a foreign body in the patient's eye.The surgeon saw the foreign body before she put the phaco tip inside the patient's eye, as she was adjusting the positioning of the sleeve and tip.The surgeon managed to retrieve the foreign body from the patient's eye.The surgery was not prolonged and no additional anesthesia was required.The patient is ok and has recovered.
|
The particle was returned for evaluation.Visual inspection found the particle was approximately 1141 x 565microns in size, white colored and hard to the touch.Analysis of the particle was performed using an energy-dispersive spectrometer (eds) equipped scanning electron microscope (sem).Eds analysis indicated the particle consisted of primarily carbon and oxygen.Lesser concentrations of chlorine, potassium, sodium, phosphorus, sulfur, calcium, and magnesium were also detected.This is consistent with organic material, saline residue and mineral deposits.Further analysis was performed using a fourier transform infrared (ftir) spectrometer.The ftir analysis of the particle produced a spectrum consistent with that of protein.Protein is not utilized in the manufacture of the product and therefore the source of the particle can not be determined.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action required.The investigation is complete.
|