Catalog Number A1059 |
Device Problems
Loose or Intermittent Connection (1371); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
This is 1 of 2 reports linked to mfg report number 3004608878-2021-00282.A facility reported that the locking knob on the mayfield modified skull clamp ((b)(4)) rotates too easily and unlocks accidentally just by lightly pushing it.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
|
|
Manufacturer Narrative
|
Mayfield modified skull clamp (a1059) was returned for evaluation.The reported complaint was confirmed via inspection of the unit.Unit received with the lock having rotational and lateral movement and requiring replacement of worn internal parts.Unit is beyond integra's 7 years recommended life cycle (unit manufactured in 2006).General maintenance, cleaning , replacement of worn components, and addition of heli-coils to the large starburst threads required at this time.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|
Event Description
|
N/a.
|
|