INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problems Loose or Intermittent Connection (1371); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
This is 1 of 2 reports linked to mfg report number 3004608878-2021-00282.A facility reported that the locking knob on the mayfield modified skull clamp ((b)(4)) rotates too easily and unlocks accidentally just by lightly pushing it.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
 
Manufacturer Narrative
Mayfield modified skull clamp (a1059) was returned for evaluation.The reported complaint was confirmed via inspection of the unit.Unit received with the lock having rotational and lateral movement and requiring replacement of worn internal parts.Unit is beyond integra's 7 years recommended life cycle (unit manufactured in 2006).General maintenance, cleaning , replacement of worn components, and addition of heli-coils to the large starburst threads required at this time.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
MDR Report Key11699576
MDR Text Key246458116
Report Number3004608878-2021-00281
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Device Lot Number064
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2021
Date Manufacturer Received05/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-