• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS PROLACTIN ASSAY; RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS PROLACTIN ASSAY; RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN) Back to Search Results
Catalog Number 07027737190
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/14/2021
Event Type  malfunction  
Manufacturer Narrative
This investigation is ongoing.This event occurred in (b)(6).Unique identifier (udi) #: (b)(4).
 
Event Description
The initial reporter received a questionable elecsys prolactin assay result for one patient with the cobas e 801 module serial number (b)(4) compared to a chemiluminescence method.Note: the units of measure were not provided for the following results.(b)(6).The questionable result was reported outside of the laboratory.It is unclear which result is deemed correct and whether the measurements were performed on the same sample.
 
Manufacturer Narrative
The calibration and qc data were requested but not provided.The investigation found the customer did not perform a polyethylene glycol precipitation (peg) for implausible high samples.Product labeling states: in case of implausible high prolactin values a precipitation by polyethylene glycol (peg) is recommended in order to estimate the amount of the biological active monomeric prolactin.The investigation determined the discrepancy is due to not following recommended procedures.This is considered improper pre-analytical handling by not performing the peg treatment.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS PROLACTIN ASSAY
Type of Device
RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key11704122
MDR Text Key269789090
Report Number1823260-2021-01204
Device Sequence Number1
Product Code CFT
Combination Product (y/n)N
PMA/PMN Number
K964748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 05/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Catalogue Number07027737190
Device Lot Number482648
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age46 YR
-
-